A study of the effect of the femoral head diameter on prosthetic hip joint dislocation using a hip-joint motion simulator
A study of the effect of the femoral head diameter on prosthetic hip joint dislocation using a hip-joint motion simulator
复制标题
利用髋关节运动模拟器研究股骨头直径对假体髋关节脱位的影响
DOI:
10.1109/iembs.2009.5334119
复制
发表时间:
2009
期刊:
影响因子:
--
通讯作者:
T. Hotokebuchi
中科院分区:
文献类型:
--
作者:
K. Kiguchi;Toru Horie;A. Yamashita;M. Ueno;Tsuneyuki Kobayashi;M. Mawatari;T. Hotokebuchi
Prosthetic hip joint dislocation phenomenon is generated by the hip joint simulator and the effect of the femoral head diameter is studied under the same condition which is similar to the condition in daily activities. Impedance control is applied to control the hip joint motion and the joint contact force simultaneously in order to take into account the constraint caused by surrounding tendons, muscles, and a joint capsule of the hip joint. The experimental results show that the hip joint dislocation phenomenon is generated by the simulator and the obtained results about the effect of the femoral head size agree with those of the other research.