How Reliable Are Enantiomeric Excess Measurements Obtained By Chiral HPLC?
How Reliable Are Enantiomeric Excess Measurements Obtained By Chiral HPLC?
复制标题
DOI:
10.1002/slct.202000166
复制
发表时间:
2020-02-14
期刊:
影响因子:
2.1
通讯作者:
Kass, Steven R.
中科院分区:
文献类型:
--
作者:
Payne, Curtis;Kass, Steven R.
Uncertainties and errors in enantiomeric excess determinations obtained by high performance liquid chromatography (HPLC) with a chiral column are evaluated with different peak resolutions and elution orders using (R)- and (S)-ibuprofen standards. Five injections of each sample were made to reduce the variability in the data and each chromatogram was monitored at multiple wavelengths to facilitate the detection of chemical impurities. Using appropriate sample concentrations, the most favorable detection wavelengths, and default instrument parameters led to large errors (>10 %) whereas properly set integration values gave impressive reproducibilities (+/- 0.5 % or better) and accuracies (