AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology
AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology
复制标题
DOI:
10.1118/1.1374247
复制
发表时间:
2001-06-01
期刊:
影响因子:
3.8
通讯作者:
Seuntjens, JP
中科院分区:
文献类型:
--
作者:
Ma, CM;Coffey, CW;Seuntjens, JP
The American Association of Physicists in Medicine (AAPM) presents a new protocol, developed by the Radiation Therapy Committee Task Group 61, fur reference dosimetry of low- and medium-energy x rays for radiotherapy and radiobiology (40kV less than or equal to tube potential less than or equal to 300 kV). It is based on ionization chambers calibrated in air in terms of air kerma. If the point of interest is at or close to the surface, one unified approach over the entire energy range shall be used to determine absorbed dose to water at the surface of a water phantom based on an in-air measurement (the "in-air" method), If the point of interest is at a depth, an in-water measurement at a depth of 2 cm shall be used for tube potentials greater than or equal to 100 kV (the "in-phantom'' method). The in-phantom method is not recommended for tube potentials