Study on absolute calibration methods for mixer-based NVNA

Study on absolute calibration methods for mixer-based NVNA
复制标题

DOI:
10.1109/cpem.2010.5544818
复制
发表时间:
2010-06
期刊:
CPEM 2010
影响因子:
--
通讯作者:
Yichi Zhang;Maoliu Lin;Zhe Zhang
Yichi Zhang;Maoliu Lin;Zhe Zhang
中科院分区:
其他
文献类型:
--
作者:
Yichi Zhang;Maoliu Lin;Zhe Zhang

文献摘要

被引文献

相似文献

本文研究了0.1 ~ 20 GHz混频器型同轴二端口非线性矢量网络分析仪的校准问题,重点研究了绝对校准技术。从8项误差模型出发,提出了两种可供选择的双端口校准方法,并通过实验验证了其结果与传统的“未知通”法相当。新方法不仅放宽了对“直通”线的要求,为标定提供了更多的选择和方便,而且为检验NVNA八项误差模型的合理性和有效性提供了一种方法。
This paper studies on the calibration of a 0.1–20GHz mixer-based two-port coaxial Nonlinear Vector Network Analyzer (NVNA), especially the absolute calibration techniques. And from the 8-term error model, we present two alternative two-port calibrating procedures, whose outcomes are revealed equivalent to those of the traditional “unknown thru” method by experiments. The new methods not only relax the requirements of “thru” line, provide more choices and more convenience for calibration, but also offer a way to examine the rationality and validity of the 8-term error model for NVNA.