Development of fast (sub‐minute) micro‐tomography
Development of fast (sub‐minute) micro‐tomography
复制标题
快速(亚分钟)显微断层扫描的发展
DOI:
10.1063/1.3478197
复制
发表时间:
2010
期刊:
影响因子:
--
通讯作者:
T. Nakano
中科院分区:
文献类型:
--
作者:
K. Uesugi;M. Hoshino;A. Takeuchi;Yoshio Suzuki;N. Yagi;T. Nakano
An x‐ray fast micro‐tomography system was developed at medical and imaging beamline BL20B2 in SPring‐8. We attempted a “continuous rotation method” whereby the sample keeps rotating in the beam during the measurement. The image acquisitions are synchronized with the trigger pulses branched from the stepper motor controller. An electron multiplier (EM)‐CCD camera (C9100‐02, 1000×1000 pixels, 30 fps, 8 μm/pixel, Hamamatsu Photonics) and beam‐monitor AA40P (Hamamatsu Photonics) were used as an image detector. The effective pixel size of the detector was 4.9 μm/pixel and the spatial resolution was about 15 μm. The total measurement time for 900 projections has been reduced to 36 s. The reconstruction time was also reduced using GPGPU and SSD system. Using these techniques, preliminary experiment for 4‐D (time resolved 3‐D) micro‐imaging has also been performed.