Analysis of 8-hydroxy-2′-deoxyguanosine in urine using high-performance liquid chromatography-electrospray tandem mass spectrometry
Analysis of 8-hydroxy-2′-deoxyguanosine in urine using high-performance liquid chromatography-electrospray tandem mass spectrometry
复制标题
DOI:
10.1016/j.jchromb.2007.03.016
复制
发表时间:
2007-06-15
影响因子:
3
通讯作者:
Peltonen, Kimmo
中科院分区:
文献类型:
--
作者:
Harri, Minna;Kasai, Hiroshi;Peltonen, Kimmo
8-Hydroxy-2'-deoxyguanosine (8-OHdG) is a widely used biomarker of oxidative stress in research related to DNA, protein damage as well as lipid peroxidation. HPLC-MS/MS with electrospray ionization (ESI) and the use of isotopically labelled 8-OHdG as an internal standard allows a simple quantification of 8-OHdG in urine samples. HPLC separation utilized the peak cutting technique and a 1.5 mm x 120 mm analytical anion exchange column. Novel method entails only minimal sample handling including the addition of a buffer and an internal standard followed by centrifugation before the samples are ready for analysis. The levels of 8-OHAG in human urine samples (n = 246) varied from 0. 16 to 16.48 mu g/L and the corresponding creatinine-normalized values were ranged from 0.49 to 14.27 mu g of 8-OHdG/g creatinine. The correlation between the developed HPLC-MS/MS method and the existing HPLC-EC method was good with an R-2 value of 0.8707. (c) 2007 Elsevier B.V. All rights reserved.