Amperometric enzyme electrode for glucose determination based on poly(pyrrole-2-aminobenzoic acid)
Amperometric enzyme electrode for glucose determination based on poly(pyrrole-2-aminobenzoic acid)
复制标题
DOI:
10.1007/s10008-009-0895-1
复制
发表时间:
2010-06
影响因子:
2.5
通讯作者:
A. Berkkan;Ali İhsan Seçkin;K. Pekmez;U. Tamer
中科院分区:
文献类型:
--
作者:
A. Berkkan;Ali İhsan Seçkin;K. Pekmez;U. Tamer
Polymerization of pyrrole and 2-aminobenzoic acid has been investigated, and a functionalized stable film of poly(pyrrole-2-aminobenzoic acid) (PP2ABA) has been obtained electrochemically onto platinum electrode. Different cyclic voltammetric behavior is obtained for polypyrrole and PP2ABA during electrosynthesis. Fourier-transformed infrared spectrometry and surface-enhanced Raman spectrometry measurements on the two films have confirmed the presence of carboxylate group in the films. The enzyme, glucose oxidase, was covalently immobilized on a conducting PP2ABA film, and amperometric response was measured as a function of concentration of glucose at a potential of 0.7 V vs Ag/AgCl in 0.1 M phosphate buffer at pH 6.2. The effect of polymeric film thickness, pH, and possible interferents were investigated. The linear range of the calibration curve is from 3 to 40 mM with a sensitivity of 0.058 µA mM−1cm−2and a limit of detection of 0.5 mM. The apperent Mishaelis–Menten constantKMis calculated to be 1 × 10−2mM, and the response time is 5 s.