Central intraluminal saturation stripe on MR angiograms of curved vessels: simulation, phantom, and clinical analysis.
Central intraluminal saturation stripe on MR angiograms of curved vessels: simulation, phantom, and clinical analysis.
复制标题
弯曲血管 MR 血管造影上的中央管腔内饱和条纹:模拟、模型和临床分析。
DOI:
10.1148/radiology.198.3.8628862
复制
发表时间:
1996
期刊:
影响因子:
--
通讯作者:
Berger,SA
中科院分区:
文献类型:
--
作者:
Saloner,D;vanTyen,R;Dillon,WP;Jou,LD;Berger,SA
PURPOSETo investigate the appearance of reduced signal intensity in the center of blood vessels on magnetic resonance (MR) angiograms that can mimic intraluminal thrombus.MATERIALS AND METHODSSimulations and phantom studies were performed to analyze MR angiogram appearance distal to a pronounced curve.RESULTSSaturation effects substantially lower the signal strength in the center of the vessel relative to that at the vessel periphery. These effects appeared even though the flow was well ordered and laminar. In curved geometries, secondary flow patterns produced counter-rotating vortices, which moved the fastest-moving particles to the outside of the curve and folded the slow-moving particles to the center of the vessel.CONCLUSIONImaging parameter choices that reduce saturation, such as acquisition of a two-dimensional section transverse to the vessel and through the questionable region, effectively eliminate the central hypointensity effect in vivo.