Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine.
Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine.
复制标题
植入心脏起搏器的放射肿瘤患者的管理:AAPM 工作组第 34 号报告。美国医学物理学家协会。
DOI:
--
复制
发表时间:
1994
期刊:
影响因子:
--
通讯作者:
A. Wolbarst
中科院分区:
文献类型:
--
作者:
J. Marbach;M. Sontag;J. V. Dyk;A. Wolbarst
Contemporary cardiac pacemakers can fail from radiation damage at doses as low as 10 gray and can exhibit functional changes at doses as low as 2 gray. A review and discussion of this potential problem is presented and a protocol is offered that suggests that radiation therapy patients with implanted pacemakers be planned so as to limit accumulated dose to the pacemaker to 2 gray. Although certain levels and types of electromagnetic interference can cause pacemaker malfunction, there is evidence that this is not a serious problem around most contemporary radiation therapy equipment.