Eight- to Ten-Year Clinical and Radiographic Outcome of a Porous Tantalum Monoblock Acetabular Component
Eight- to Ten-Year Clinical and Radiographic Outcome of a Porous Tantalum Monoblock Acetabular Component
复制标题
DOI:
10.1016/j.arth.2008.06.020
复制
发表时间:
2009-08-01
影响因子:
3.5
通讯作者:
Papagelopoulos, Panayiotis J.
中科院分区:
文献类型:
--
作者:
Macheras, George;Kateros, Konstantinos;Papagelopoulos, Panayiotis J.
In a prospective study, the authors used a porous tantalum monoblock acetabular component for primary total hip arthroplasty between November 1997 and June 1999. A total of 156 consecutive primary total hip arthroplasty were done in 143 patients younger than 75 years. A total of 151 hips had a follow-up time from 8 to 10 years. The average preoperative total Harris hip score of 44.0 +/- 13.8 increased to 97.0 +/- 6.2 at the latest follow-Lip. The average preoperative Oxford hip score of 43.3 +/- 6.5 improved to 13.9 +/- 2.3 at the latest follow-up. Radiographic evaluation including the Ein-Bild-Rontgen-Analyse (EBRA) digital system showed no radiographic evidence of gross polyethylene wear, progressive radiolucencies, osteolytic lesions, acetabular fracture, or component Subsidence. There were 7 (4.5%) postoperative complications all unrelated to the acetabular component.