Detection of femtomolar concentrations of HF using an SiO2 microcantilever
Detection of femtomolar concentrations of HF using an SiO2 microcantilever
复制标题
DOI:
10.1021/ac035140g
复制
发表时间:
2004-05-01
影响因子:
7.4
通讯作者:
Thundat, T
中科院分区:
文献类型:
--
作者:
Tang, YJ;Fang, J;Thundat, T
Femtomolar concentrations of hydrogen fluoride, a decomposition component of nerve agents, were detected using a SiO2 microcantilever. The microcantilever underwent bending due to the reaction of HF with SiO2. The microcantilever deflection increased as the concentration of HF increased. Other acids, such as HCl, had no effect on the deflection of the cantilever. The mechanism of reaction-induced bending and the correlation of microcantilever deflection with the HF concentration are discussed. The deflection in response to HF of a commercially available silicon cantilever was also studied, and its response was compared with that of the SiO2 cantilever. Much less bending amplitude and sensitivity were observed for the silicon cantilever.