In-bore prostate transperineal interventions with an MRI-guided parallel manipulator: system development and preliminary evaluation.
In-bore prostate transperineal interventions with an MRI-guided parallel manipulator: system development and preliminary evaluation.
复制标题
DOI:
10.1002/rcs.1671
复制
发表时间:
2016-06
期刊:
影响因子:
--
通讯作者:
Iordachita I
中科院分区:
文献类型:
--
作者:
Eslami S;Shang W;Li G;Patel N;Fischer GS;Tokuda J;Hata N;Tempany CM;Iordachita I
The robot-assisted minimally-invasive surgery is well recognized as a feasible solution for diagnosis and treatment of the prostate cancer in human. In this paper the kinematics of a parallel 4 Degrees-of-Freedom (DOF) surgical manipulator designed for minimally invasive in-bore prostate percutaneous interventions through the patient's perineum. The proposed manipulator takes advantage of 4 sliders actuated by MRI-compatible piezoelectric motors and incremental rotary encoders. Errors, mostly originating from the design and manufacturing process, need to be identified and reduced before the robot is deployed in the clinical trials. The manipulator has undergone several experiments to evaluate the repeatability and accuracy of the needle placement which is an essential concern in percutaneous prostate interventions. The acquired results endorse the sustainability, precision (about 1 mm in air (in x or y direction) at the needle's reference point) and reliability of the manipulator.