Monitoring temporary immunodepression by flow cytometric measurement of monocytic HLA-DR expression:: A multicenter standardized study
Monitoring temporary immunodepression by flow cytometric measurement of monocytic HLA-DR expression:: A multicenter standardized study
复制标题
DOI:
10.1373/clinchem.2005.052639
复制
发表时间:
2005-12-01
影响因子:
9.3
通讯作者:
Kunz, D
中科院分区:
文献类型:
--
作者:
Döcke, WD;Höflich, C;Kunz, D
Background: Single-center trials have shown that monocytic HLA-DR is a good marker for monitoring the severity of temporary immunodepression after trauma, major surgery, or sepsis. A new test for measuring monocytic HLA-DR is now available.Methods: We evaluated a new test reagent set for monocytic HLA-DR expression (BD Quantibrite (TM) HLA-DR/Monocyte reagent; Becton Dickinson) in single-laboratory and interlaboratory experiments, assessing preanalytical handling, lyse-no-wash (LNW) vs lyse-wash (LW) values, reference values, and the effect of use of different flow cytometers and different instrument settings on test variance.Results: For preanalytical handling, EDTA anticoagulation, storage on ice as soon as possible, and staining within 4 h after blood collection gave results comparable to values obtained for samples analyzed immediately after collection (mean increase of similar to 4% in monocytic HLA-DR). Comparison of LNW and LW revealed slightly higher results for LNW (similar to 18% higher for LNW compared with LW; r = 0.982). Comparison of different flow cytometers and instrument settings gave CVs