Methods to calibrate the absolute receive sensitivity of single-element, focused transducers.
Methods to calibrate the absolute receive sensitivity of single-element, focused transducers.
复制标题
DOI:
10.1121/1.4929620
复制
发表时间:
2015-09
期刊:
影响因子:
--
通讯作者:
Kyle T. Rich;T. D. Mast
中科院分区:
文献类型:
--
作者:
Kyle T. Rich;T. D. Mast
Absolute pressure measurements of acoustic emissions by single-element, focused passive cavitation detectors would be facilitated by improved wideband receive calibration techniques. Here, calibration methods were developed to characterize the absolute, frequency-dependent receive sensitivity of a spherically focused, single-element transducer using pulse-echo and pitch-catch techniques. Validation of these calibration methods on a focused receiver were made by generating a pulse from a small diameter source at the focus of the transducer and comparing the absolute pressure measured by a calibrated hydrophone to that of the focused transducer using the receive sensitivities determined here.