Marker-free method for accurate alignment between correlated light, cryo-light, and electron cryo-microscopy data using sample support features
Marker-free method for accurate alignment between correlated light, cryo-light, and electron cryo-microscopy data using sample support features
复制标题
DOI:
10.1016/j.jsb.2017.11.001
复制
发表时间:
2018-01-01
影响因子:
3
通讯作者:
Volkmann, Niels
中科院分区:
文献类型:
--
作者:
Anderson, Karen L.;Page, Christopher;Volkmann, Niels
Combining fluorescence microscopy with electron cryo-tomography allows, in principle, spatial localization of tagged macromolecular assemblies and structural features within the cellular environment. To allow precise localization and scale integration between the two disparate imaging modalities, accurate alignment procedures are needed. Here, we describe a marker-free method for aligning images from light or cryo-light fluorescence microscopy and from electron cryo-microscopy that takes advantage of sample support features, namely the holes in the carbon film. We find that the accuracy of this method, as judged by prediction errors of the hole center coordinates, is better than 100 nm.