Aorto-iliac multidetector-row CT angiography with low kV settings: improved vessel enhancement and simultaneous reduction of radiation dose
Aorto-iliac multidetector-row CT angiography with low kV settings: improved vessel enhancement and simultaneous reduction of radiation dose
复制标题
DOI:
10.1007/s00330-004-2575-y
复制
发表时间:
2005-02-01
影响因子:
5.9
通讯作者:
Becker, C
中科院分区:
文献类型:
--
作者:
Wintersperger, B;Jacobs, T;Becker, C
The aim of the study was to implement an abdominal CT angiography protocol using 100 kVp and to compare SNR and CNR, as well as subjective image quality, to a standard CT angiography protocol using 120 kVp on a 16 detector-row CT scanner. Forty-eight patients were referred for routine abdominal CT angiography on a 16 detector-row CT scanner. Patients were scanned using either 120 or 100 kVp at constant mAs settings. Vessel opacification was provided by automated contrast injection using similar injection protocols. Density measurements were performed along the aorto-iliac axis with SNR and CNR calculation. In addition, the estimated effective patient radiation dose was calculated. Results of both protocols were compared. The 100-kVp protocol (432180 HU) showed a significantly higher vessel density than the 120-kVp (333 /- 90 HU; P