Diffusion coefficient measurement using a temperature-controlled fluid for quality control in multicenter studies.

Diffusion coefficient measurement using a temperature-controlled fluid for quality control in multicenter studies.
复制标题

DOI:
10.1002/jmri.22363
复制
发表时间:
2011-10
影响因子:
4.4
通讯作者:
Ross, Brian D.
Ross, Brian D.
中科院分区:
医学2区
文献类型:
--
作者:
Chenevert, Thomas L.;Galban, Craig J.;Ivancevic, Marko K.;Rohrer, Susan E.;Londy, Frank J.;Kwee, Thomas C.;Meyer, Charles R.;Johnson, Timothy D.;Rehemtulla, Alnawaz;Ross, Brian D.

文献摘要

参考文献

被引文献

相似文献

介绍一种DW-MRI质量控制冰水体模的应用。DW-MRI已成为一种重要的肿瘤影像生物标志物,可用于诊断和早期治疗反应评估。通过多中心试验验证成像生物标记物需要跨地点的校准和性能测试。这个模子由一个中心管组成,里面装满了蒸馏水,周围环绕着冰水。在制备模体后,让大约30分钟达到热平衡。DW-MRI数据来自7个机构,来自3个供应商的20台MRI扫描仪和2个场强(1.5和3T)。在25天的时间里,在16天的不同时间里,该模型还在一个单一系统上进行了扫描。所有数据都被转移到密歇根大学的中央处理站点进行分析。结果表明,所有被测系统之间的测量ADC值的偏差为±5%,表明系统之间具有很好的一致性。单一系统在25天内的重复性也被发现在±5%的ADC值内。总体而言,在多中心试验中,使用冰水模体评估ADC被发现是一个合理的候选方案。这里描述的冰水体模是一种实用而通用的方法,可以随着MRI序列和硬件设计的不断变化来验证ADC测量的准确性,并且可以很容易地在多中心临床试验设计中实施。
To present the use of a quality control ice-water phantom for DW-MRI. DW-MRI has emerged as an important cancer imaging biomarker candidate for diagnosis and early treatment response assessment. Validating imaging biomarkers through multi-center trials requires calibration and performance testing across sites. The phantom consisted of a center tube filled with distilled water surrounded by ice-water. Following preparation of the phantom approximately 30 minutes was allowed to reach thermal equilibrium. DW-MRI data was collected at 7 institutions, 20 MRI scanners from three vendors and 2 field strengths (1.5 and 3T). The phantom was also scanned on a single system on 16 different days over a 25 day period. All data was transferred to a central processing site at the University of Michigan for analysis. Results revealed that the variation of measured ADC values between all systems tested was ±5% indicating excellent agreement between systems. Reproducibility of a single system over a 25 day period was also found to be within ±5% ADC values. Overall, the use of an ice water phantom for assessment of ADC was found to be a reasonable candidate for use in multi-center trials. The ice water phantom described here is a practical and universal approach to validate the accuracy of ADC measurements with ever changing MRI sequence and hardware design and can be readily implemented in multicenter clinical trial designs.
DOI: 10.1016/j.jmr.2007.07.003
发表时间: 2007-11-01
影响因子: 2.2
作者:
Komlosh, M. E.;Horkay, F.;Basser, P. J.
通讯作者: Basser, P. J.
DOI: 10.1593/tlo.09175
发表时间: 2009-09-01
影响因子: 5
作者:
Galban, Craig J.;Mukherji, Suresh K.;Ross, Brian D.
通讯作者: Ross, Brian D.
DOI: 10.1039/b005319h
发表时间: 2000-01-01
影响因子: 3.3
作者:
Holz, M;Heil, SR;Sacco, A
通讯作者: Sacco, A
DOI: 10.1002/nbm.1328
发表时间: 2009-04-01
期刊: NMR IN BIOMEDICINE
影响因子: 2.9
作者:
Riches, S. F.;Hawtin, K.;de Souza, N. M.
通讯作者: de Souza, N. M.
DOI: 10.1016/j.nic.2009.08.009
发表时间: 2009-11
影响因子: 2.3
作者:
Chenevert, Thomas L.;Ross, Brian D.
通讯作者: Ross, Brian D.