Comparison of low-temperature hydrogen peroxide gas plasma sterilization for endoscopes using various Sterrad™ models

Comparison of low-temperature hydrogen peroxide gas plasma sterilization for endoscopes using various Sterrad™ models
复制标题

DOI:
10.1016/j.jhin.2004.10.002
复制
发表时间:
2005-04-01
影响因子:
6.9
通讯作者:
Borneff-Lipp, M
Borneff-Lipp, M
中科院分区:
医学3区
文献类型:
--
作者:
Okpara-Hofmann, J;Knoll, M;Borneff-Lipp, M

文献摘要

被引文献

相似文献

这项研究比较了不同型号的STERRAD(TM)系统(STERRAD 50、100、100S和200)对四种类型内窥镜的灭菌效果。所有接种了嗜热脂肪杆菌孢子的携带者都达到了符合国际要求的灭菌水平。内窥镜的测试周期为半个周期(过度杀伤力)。这是第一次对目前市场上销售的STERRAD模型在窄腔内窥镜的有效灭菌方面进行比较的研究。(C)2004年医院感染学会。爱思唯尔有限公司出版。保留所有权利。
This study compared the effectiveness of sterilizing four types of endoscope using different models of the Sterrad (TM) system (Sterrad 50, 100, 100S and 200). Sterilization levels meeting international requirements were attained in all cases with carriers inoculated with Geobacillus stearothermophilus spores. The endoscopes were tested in half cycles ('overkill'). This is the first study to compare the Sterrad models marketed to date in terms of effective sterilization of endoscopes with narrow lumens. (c) 2004 The Hospital Infection Society. Published by Elsevier Ltd. All rights reserved.