RECOMMENDATIONS FOR STANDARDIZATION AND SPECIFICATIONS IN AUTOMATED ELECTROCARDIOGRAPHY - BANDWIDTH AND DIGITAL SIGNAL-PROCESSING - A REPORT FOR HEALTH-PROFESSIONALS BY AN AD HOC WRITING GROUP OF THE COMMITTEE ON ELECTROCARDIOGRAPHY AND CARDIAC ELECTROPHYSIOLOGY OF THE COUNCIL-ON-CLINICAL-CARDIOLOGY, AMERICAN-HEART-ASSOCIATION
RECOMMENDATIONS FOR STANDARDIZATION AND SPECIFICATIONS IN AUTOMATED ELECTROCARDIOGRAPHY - BANDWIDTH AND DIGITAL SIGNAL-PROCESSING - A REPORT FOR HEALTH-PROFESSIONALS BY AN AD HOC WRITING GROUP OF THE COMMITTEE ON ELECTROCARDIOGRAPHY AND CARDIAC ELECTROPHYSIOLOGY OF THE COUNCIL-ON-CLINICAL-CARDIOLOGY, AMERICAN-HEART-ASSOCIATION
复制标题
DOI:
10.1161/01.cir.81.2.730
复制
发表时间:
1990-02-01
期刊:
影响因子:
37.8
通讯作者:
ZYWIETZ, C
中科院分区:
文献类型:
--
作者:
BAILEY, JJ;BERSON, AS;ZYWIETZ, C
R ecent technologic advancesand published studies point to the need to reevaluate requirements for recording and analysis of electrocardiographic (ECG) data. Severalissues have arisen from recent developments. First, approximately 52 million ECGs were processed by computer in 1987 (up from 4 million in 1975), and more than 15,000 digital ECGsystems are now in use (up from 85 in 1975). 1 In addition, computer-assisted electrocardiography is spreading rapidly from academic medical centers to private hospitals, clinics, and physicians' offices. A major factor in this diffusion of technology is the introduction of inexpensive (less than $5,000) stand-alone interpretative carts using microchip technology.'Second, a large number of ECG systems, particu-larly stand-alone carts, convert data into digital form immediately after amplification andprocess it digitally. This type of processing, combined with pream-plifiers with lower noise characteristics than those available earlier, permitshigher resolutions and signal-to-noise ratios than conventional linear analog systems.