The impact of motion artifacts on the reproducibility of repeated coronary artery calcium measurements
The impact of motion artifacts on the reproducibility of repeated coronary artery calcium measurements
复制标题
DOI:
10.1007/s00330-006-0278-2
复制
发表时间:
2007-01-01
影响因子:
5.9
通讯作者:
Ito, Katsuhide
中科院分区:
文献类型:
--
作者:
Horiguchi, Jun;Fukuda, Hiroshi;Ito, Katsuhide
The purpose of this study is, using a 16-section multidetector-row helical computed tomography ( MDCT) scanner with retrospective reconstruction, to compare variability in repeated coronary calcium scoring and qualitative scores of the motion artifacts. One hundred forty-four patients underwent two subsequent scans using MDCT. According to Agatston and volume algorithms, the coronary calcium scores during mid-diastole ( the center corresponding to 70% of the R-R cycle) were calculated and the inter-scan variability was obtained. Motion artifacts from coronary artery calcium were subjectively evaluated and classified using a 5-point scale: 1, excellent; no motion artifacts; 2, fine, minor motion artifacts; 3, moderate, mild motion artifacts; 4, bad, severe motion artifacts; 5, poor, doubling or discontinuity. Each reading was done by vessels ( left main, left descending, left circumflex and right coronary arteries) and the motion artifact score ( mean of the scales) was determined per patient. The variability in the low (1.2 +/- 0.2) and high (2.4 +/- 0.6) motion artifact score groups was 7 +/- 6 ( median, 6)% and 19 +/- 15 ( 16)% on the Agatston score (P< 0.01) and 7 +/- 7 ( 6)% and 16 +/- 13 ( 14)% on the volume score ( P< 0.01), respectively. In conclusion, motion has a significant impact on the reproducibility of coronary calcium scoring.