Finite element model of the temperature increase in excised porcine cadaver iris during direct illumination by femtosecond laser pulses.
Finite element model of the temperature increase in excised porcine cadaver iris during direct illumination by femtosecond laser pulses.
复制标题
飞秒激光脉冲直接照射期间离体猪尸体虹膜温度升高的有限元模型。
DOI:
10.1117/1.jbo.17.7.078001
复制
发表时间:
2012
影响因子:
3.5
通讯作者:
Juhasz,Tibor
中科院分区:
文献类型:
--
作者:
Sun,Hui;Kurtz,RonaldM;Juhasz,Tibor
In order to model the thermal effect of laser exposure of the iris during laser corneal surgery, we simulated the temperature increase in porcine cadaver iris. The simulation data for the 60 kHz FS60 Laser showed that the temperature increased up to 1.23°C and 2.45°C (at laser pulse energy 1 and 2µJ, respectively) by the 24 second procedure time. Calculated temperature profiles show good agreement with data obtained fromex vivoexperiments using porcine cadaver iris. Simulation results of different types of femtosecond lasers indicate that the Laser in situ keratomileusis procedure does not present a safety hazard to the iris.