A compact imaging system with a CdTe double-sided strip detector for non-destructive analysis using negative muonic X-rays
A compact imaging system with a CdTe double-sided strip detector for non-destructive analysis using negative muonic X-rays
复制标题
DOI:
10.1016/j.nima.2017.11.004
复制
发表时间:
2018-12-21
影响因子:
1.4
通讯作者:
Yabu, Goro
中科院分区:
文献类型:
--
作者:
Katsuragawa, Miho;Tampo, Motonobu;Yabu, Goro
A CdTe double-sided strip detector (CdTe-DSD) is an ideal device for imaging and spectroscopic measurements in the hard X-ray range above 10 keV. Recent development enables us to realize an imager with a detection area of similar to 10 cm(2). An energy resolution of 1-2 keV (FWHM) and a position resolution of a few hundred mu m are available from the detector. This type of imager has been long awaited for non-destructive elemental analysis, especially by using negative muons, because energies of characteristic X-rays from muonic atoms are about 200 time higher than those from normal atoms. With the method that uses negative muons, hard X-ray information gives the spatial distribution of elements in samples at a certain depth defined by the initial momentum of the muon beam. In order to study three-dimensional imaging capability of the method, we have developed a compact imaging system based on CdTe-DSD and a phi 3 mm pinhole collimator as the first prototype. We conducted experiments with samples which consist of layers of Al, BN and LiF irradiated by negative muon beams in MUSE/J-PARC and successfully reconstruct hard X-ray images of muonic X-rays from B, N and F at various depths. (C) 2017 Elsevier B.V. All rights reserved.