Design of a breast phantom for quantitative MRI
Design of a breast phantom for quantitative MRI
复制标题
DOI:
10.1002/jmri.25214
复制
发表时间:
2016-09-01
影响因子:
4.4
通讯作者:
Hylton, Nola M.
中科院分区:
文献类型:
--
作者:
Keenan, Kathryn E.;Wilmes, Lisa J.;Hylton, Nola M.
PurposeWe present a breast phantom designed to enable quantitative assessment of measurements of T-1 relaxation time, apparent diffusion coefficient (ADC), and other attributes of breast tissue, with long-term support from a national metrology institute.Materials and MethodsA breast phantom was created with two independent, interchangeable units for diffusion and T-1/T-2 relaxation, each with flexible outer shells. The T-1 unit was filled with corn syrup solution and grapeseed oil to mimic the relaxation behavior of fibroglandular and fatty tissues, respectively. The diffusion unit contains plastic tubes filled with aqueous solutions of polyvinylpyrrolidone (PVP) to modulate the ADC. The phantom was imaged at 1.5T and 3.0T using magnetic resonance imaging (MRI) scanners and common breast coils from multiple manufacturers to assess T-1 and T-2 relaxation time and ADC values.ResultsThe fibroglandular mimic exhibited target T-1 values on 1.5T and 3.0T clinical systems (25-75 percentile range: 1289 to 1400 msec and 1533 to 1845 msec, respectively) across all bore temperatures. PVP solutions mimicked the range of ADC values from malignant tumors to normal breast tissue (40% PVP median: 633 x 10(-6) mm(2)/s to 0% PVP median: 2231 x 10(-6) mm(2)/s) at temperatures of 17-24 degrees C. The interchangeable phantom units allowed both the diffusion and T-1/T-2 units to be tested on the left and right sides of the coil to assess any variation.ConclusionThis phantom enables T-1 and ADC measurements, fits in a variety of clinical breast coils, and can serve as a quality control tool to facilitate the standardization of quantitative measurements for breast MRI. J. Magn. Reson. Imaging 2016;44:610-619.