Comparative Assessment of Anti-HLA Antibodies Using Two Commercially Available Luminex-Based Assays
Comparative Assessment of Anti-HLA Antibodies Using Two Commercially Available Luminex-Based Assays
复制标题
DOI:
10.1097/txd.0000000000000734
复制
发表时间:
2017-11-01
影响因子:
2.3
通讯作者:
Zorn, Emmanuel
中科院分区:
文献类型:
--
作者:
Clerkin, Kevin J.;See, Sarah B.;Zorn, Emmanuel
Background. Allospecific anti-HLA antibodies (Abs) are associated with rejection of solid organ grafts. The 2main kits to detect anti-HLA Ab in patient serum are commercialized by Immucor and One Lambda/ThermoFisher. We sought to compare the performance of both platforms. Methods. Background-adjusted mean fluorescence intensity (MFI) values were used from both platforms to compare sera collected from 125 pretransplant and posttransplant heart and lung transplant recipients. Results. Most HLA class I (94.5%) and HLA class II (89%) Abs with moderate to high MFI titer (>= 4000) were detected by both assays. A modest correlation was observed between MFI values obtained from the 2 assays for both class I (r = 0.3, r(2) = 0.09, P < 0.0001) and class II Ab (r = 0.707, r(2) = 0.5, P < 0.0001). Both assays detected anti-class I and II Ab that the other did not; however, no specific HLA allele was detected preferentially by either of the 2 assays. For a limited number of discrepant sera, dilution resulted in comparable reactivity profiles between the 2 platforms. Conclusions. Immucor and One Lambda/ ThermoFisher assays have a similar, albeit nonidentical, ability to detect anti-HLA Ab. Although the correlation between the assays was present, significant variances exist, some of which can be explained by a dilution-sensitive "prozone" effect.