Determination of Trace Rhenium in Geological Samples using Isotope Dilution-Inductively Coupled Plasma Mass Spectrometry

Determination of Trace Rhenium in Geological Samples using Isotope Dilution-Inductively Coupled Plasma Mass Spectrometry
复制标题

DOI:
--
复制
发表时间:
2006
期刊:
--
影响因子:
--
通讯作者:
Cheng Jiangfeng
Cheng Jiangfeng
中科院分区:
其他
文献类型:
--
作者:
Cheng Jiangfeng

文献摘要

被引文献

相似文献

开发了一种使用同位素稀释-电感耦合等离子体质谱法测定地质样品(黄铁矿和岩石)中痕量铼的方法。采用Carius管消化法进行样品制备。总程序空白可限制在 1. 4 - 10 pg 范围内。通过分析国标辉钼矿样品对方法进行了验证,结果与标准值(17. 39±0. 32)μg/g吻合较好。该方法用于岩石样品中痕量Re(1.27 ng/g)的测定,精度为3. 89% RSD(n=4, 2s)。
A method for the determination of trace rhenium in geological samples (pyrites and rocks) using isotope dilution-inductively coupled plasma mass spectrometry was developed. Carius tube digestion method was used for sample preparation. The total procedure blanks could be limited in the range of 1. 4 - 10 pg. The method was verified by analyzing the national standard reference molybdenite sample and the result was in good agreement with the certified value (17. 39±0. 32) μg/g. The method has been applied to the determination of trace Re (1.27 ng/g) in rock sample with precision of 3. 89% RSD(n=4, 2s).