Integrating sphere effective optical path length calibration by gas absorption spectroscopy
Integrating sphere effective optical path length calibration by gas absorption spectroscopy
复制标题
气体吸收光谱积分球有效光程长度校准
DOI:
--
复制
发表时间:
2014
期刊:
影响因子:
--
通讯作者:
Wei Guo
中科院分区:
文献类型:
--
作者:
Q. Gao;Yungang Zhang;Jia Yu;Zhiguo Zhang;Shaohua Wu;Wei Guo
A method of integrating sphere effective optical path length (EOPL) evaluation using tunable diode laser absorption spectroscopy for gas detection was demonstrated. Oxygen was used as a sample gas for an 8.38 cm diameter integrating sphere calibration; 393.7 ± 1.3 cm EOPL was obtained from the wavelength modulation spectroscopy with second harmonic calibration by measuring oxygen P11 line at 764 nm, which is in agreement with that of 393 cm by using direct absorption spectroscopy calibration. The EOPL calibration accuracy of this method can reach 0.33 %. It has been justified that the EOPL of an integrating sphere is independent of the incident light intensity.