Improving energy stability in the National Institute of Standards and Technology Microcalorimeter X-ray detector
Improving energy stability in the National Institute of Standards and Technology Microcalorimeter X-ray detector
复制标题
提高美国国家标准技术研究院微量热仪 X 射线探测器的能量稳定性
DOI:
10.1154/1.1913706
复制
发表时间:
2005
影响因子:
0.5
通讯作者:
D. Newbury
中科院分区:
文献类型:
--
作者:
T. Jach;J. Small;D. Newbury
Microcalorimeter X-ray detectors using a transition edge sensor depend for their linearity and energy scale on the stability of the operating point on the transition curve. We report on some sources of energy scale drift in microcalorimeter X-ray detectors and the manner in which they have been addressed. Previously observed drifts of >10 eV∕h have been reduced to 1–2 eV∕h. This improved stability has resulted in the observation of X-ray fluorescence linewidths of ⩽12 eV over 6 h of counting time.