MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE
MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE
复制标题
DOI:
10.1088/0031-9155/35/9/002
复制
发表时间:
1990-09-01
影响因子:
3.5
通讯作者:
DANCE, DR
中科院分区:
文献类型:
--
作者:
DANCE, DR
The IPSM report on the commissioning and routine testing of mammographic x-ray systems recommends that breast dose be specified as the mean dose to the glandular tissues within the breast and gives the size and compositions of a standard breast phantom for the comparison of doses. The dose to this standard breast phantom can be determined by measuring the incident air kerma to a Perspex phantom and applying appropriate multiplicative conversion factors. These conversion factors have been evaluated by Monte Carlo calculations for a wide range of mammographic x-ray spectra. Some factors are provided for a range of breast thicknesses to supplement existing tabulations. Results are also given for equivalent thickness of Perspex and breast tissue.