Development of a 35-MHz piezo-composite ultrasound array for medical imaging
Development of a 35-MHz piezo-composite ultrasound array for medical imaging
复制标题
DOI:
10.1109/tuffc.2006.1588408
复制
发表时间:
2006-01-01
影响因子:
3.6
通讯作者:
Shung, KK
中科院分区:
文献类型:
--
作者:
Cannata, JM;Williams, JA;Shung, KK
This paper discusses the development of a 64-element 35-MHz composite ultrasonic array. This array was designed primarily for ocular imaging applications, and features 2-2 composite elements mechanically diced out of a fine-grain high-density Navy Type VI ceramic. Array elements were spaced at a 50-micron pitch, interconnected via a custom flexible circuit and matched to the 50-mu m system electronics via a 75-mu m transmission line coaxial cable. Elevation focusing was achieved using a cylindrically shaped epoxy lens. One functional 64-element array was fabricated and tested. Bandwidths averaging 55%, 23-dB insertion loss, and crosstalk less than -24 dB were measured. An image of a tungsten wire target phantom was acquired using a synthetic aperture reconstruction algorithm. The results from this imaging test demonstrate resolution exceeding 50 mu m axially and 100 mu m laterally.