Characterisation of a multi-channel multiplexed EMG recording system: towards realising variable electrode configurations
Characterisation of a multi-channel multiplexed EMG recording system: towards realising variable electrode configurations
复制标题
多通道复用肌电图记录系统的表征:实现可变电极配置
DOI:
10.1109/biocas.2019.8918710
复制
发表时间:
2019
期刊:
影响因子:
--
通讯作者:
De Jager K
中科院分区:
文献类型:
--
作者:
De Jager K
An implantable EMG amplifier with a novel multiplexed frontend for in vivo selection of optimal electrode configurations, was designed using commercially available components. The multiplexers are advantageous as optimal electrode configuration are not known before implantation. The system has 6 ADC recording channels (ADS1298 biopotential amplifier, 2 kHz sampling frequency, 16-bit resolution), and three 8 × 8 multiplexer arrays (ADG2188), 2 channels per MUX. The system was characterised by measuring input impedance (5.8 MOhm) and frequency response (CMRR 49.0 dB; SNR 51.4 dB). EMG recordings from implanted epimysial electrodes showed lower signal quality (13.5 dB) compared with a commercial EMG recorder (19.5 dB), nonetheless the signals appeared suitable for myoelectric control applications. An implantable version of the EMG recorder, housed within a hermetically sealed ceramic package, should improve signal quality.