RF aitifacts caused by metallic implants or instruments which get more prominent at 3 T: an in vitro study

RF aitifacts caused by metallic implants or instruments which get more prominent at 3 T: an in vitro study
复制标题

DOI:
10.1016/j.mri.2004.12.009
复制
发表时间:
2005-04-01
影响因子:
2.5
通讯作者:
Schick, F
Schick, F
中科院分区:
医学4区
文献类型:
--
作者:
Graf, H;Lauer, UA;Schick, F

文献摘要

被引文献

相似文献

在体内或邻近具有高导电性的金属装置可能导致被调查组织中RF振幅B-1的显着改变,特别是在RF频率增加时,如果满足特定的电磁共振条件。在B-0=0.2, 1.5和3t时系统地研究了rf -金属相互作用效应,分析了铜线(d=1 mm, L=53和27 cm)以及以下钛或镍钛诺制成的器械和植入物的相关图像伪影:活检针、髋关节假体、血管支架和动脉瘤夹。样品在gd - dtpa掺杂的140 mM NaCl溶液中使用高读出带宽的自旋回波(SE)序列进行检测。采用自动变送器调节V-T、v -自动变送器电压V-T和手动降低变送器电压V-T检测B-1增强。在0.2 T时,超出支架管腔区域的屏蔽,没有观察到射频效应。在1.5 T时,铜线产生明显的射频伪影。在3t时,髋关节假体和活检针也出现射频伪影。在较低场强下具有明显腔内屏蔽的支架,在其外表面附近显示明显的B增强。©2005 Elsevier Inc.版权所有。
Metallic devices with high electrical conductivity inside or adjacent to the body might lead to marked alterations of the RF amplitude B-1 in the tissue under investigation, especially at increased RF frequency, and if specific conditions for electromagnetic resonance are fulfilled. RF-metal interaction effects were investigated systematically at B-0=0.2, 1.5 and 3 T analyzing correlated image artifacts for copper wires (d=1 mm, L=53 and 27 cm), and for following instruments and implants made of titanium or nitinol: biopsy needles, hip prostheses, vascular stents and aneurysm clips. The samples were examined in Gd-DTPA-doped 140 mM NaCl solution using spin-echo (SE) sequences with high readout bandwidth. Automatic transmitter adjustment V-T,V-auto and manually reduced transmitter voltage V-T were applied in order to detect B-1 enhancement. At 0.2 T, beyond the shielding of the luminal region of the stents, no RF effects were observed. At 1.5 T, the copper wires caused distinct RF artifacts. At 3 T, RF artifacts also appeared for the hip prostheses and the biopsy needles. Stents with pronounced luminal shielding at lower field strength revealed marked B, enhancement close to their outer surface. © 2005 Elsevier Inc. All rights reserved.