The BEPC II Luminosity Monitor (BLM) monitors relative luminosity per bunch. The counting rates of gamma photons, which are proportional to the luminosities from the BLM at the center of mass system energy of the psi(3770) resonance, are obtained with a statistical error of 0.01% and a systematic error of 4.1%. Absolute luminosities are also determined by the BESIII End-cap Electro-Magnetic Calorimeter (EEMC) using Bhabha events with a statistical error of 2.3% and a systematic error of 3.5%. The calibration constant between the luminosities obtained with the EEMC and the counting rates of the BLM are found to be 0.84 +/- 0.03 (x10(26) cm(-2).count(-1)). With the calibration constant, the counting rates of the BLM can be scaled up to absolute luminosities.