Possibility of cardiac output monitoring from the intra-arterial blood pressure profile.
Possibility of cardiac output monitoring from the intra-arterial blood pressure profile.
复制标题
通过动脉内血压曲线监测心输出量的可能性。
DOI:
10.3109/10641968509073555
复制
发表时间:
1985
期刊:
影响因子:
--
通讯作者:
G. Muiesan
中科院分区:
文献类型:
--
作者:
C. Alicandri;R. Fariello;E. Boni;A. Zaninelli;G. Muiesan
A method for estimating cardiac output (CO) from the intra-arterial blood pressure profile ("contour method") was tested in 8 patients: 6 with essential hypertension, 1 with a pheochromocytoma and 1 with orthostatic hypotension. CO (1/min) was derived by the following formula: PSA (1+St/Dt) X HR 10(-3), where PSA is the area under the systolic portion of the pressure curve, St is the systolic and Dt the diastolic time, X is a correction factor, HR is the heart rate and 10(-3) is a conversion factor from ml/min to 1/min. The "contour method" was compared to the thermodilution CO method. The correlation between the 2 methods was highly significant: the r value in all patients during different conditions (supine, tilt, dynamic and static exercise) ranged from 0.91 to 0.97 with an intercept close to 0 and a slope close to 1. These results indicate that CO is properly measured from the intra-arterial blood pressure profile by the "contour method". A continuous hemodynamic monitoring can be derived applying the "contour method" to the intra-arterial blood pressure profile obtained with the Oxford technique.