Application of a C-peptide electrospray ionization-isotope dilution-liquid chromatography-tandem mass spectrometry measurement procedure for the evaluation of five C-peptide immunoassays for urine
Application of a C-peptide electrospray ionization-isotope dilution-liquid chromatography-tandem mass spectrometry measurement procedure for the evaluation of five C-peptide immunoassays for urine
复制标题
DOI:
10.1016/s1570-0232(03)00268-x
复制
发表时间:
2003-07-25
影响因子:
3
通讯作者:
Thienpont, LM
中科院分区:
文献类型:
--
作者:
Fierens, C;Stöckl, D;Thienpont, LM
This study applied electrospray ionization-isotope dilution-liquid chromatography-tandem mass spectrometry for the evaluation of five urinary C-peptide immumoassays via split-sample measurements. The immunoassays measured in duplicate in the same run, the comparison method in triplicate over different runs. From the data, the within-run imprecision and the method comparison total RSDs were calculated. Regression analysis revealed on the one hand systematic differences, on the other. an excellent correlation between the test and comparison methods. From the spread of the data around the regression line in comparison with the 95% prediction intervals from the total RSD, sample-related effects and/or specificity problems were apparent and investigated. (C) 2003 Elsevier B.V. All rights reserved.