Technical note: Analysis of brachytherapy source movement by high-speed camera
Technical note: Analysis of brachytherapy source movement by high-speed camera
复制标题
DOI:
10.1002/mp.15601
复制
发表时间:
2022-05-13
期刊:
影响因子:
3.8
通讯作者:
Igaki, Hiroshi
中科院分区:
文献类型:
--
作者:
Okamoto, Hiroyuki;Iijima, Kotaro;Igaki, Hiroshi
Purpose To assess the accuracy of the movement of a brachytherapy source using a high-speed camera for evaluating source position, dwell time, and transit dose. Methods A high-speed camera was used to record the source position of an Ir-192 source relative to a ruler within a custom positioning jig in a remote afterloading system. The analyzed frames can be used to assess dwell positions and times. Treatment plans had multiple dwell times equal to 0.1, 0.5, 1.0, and 2.0 s in 2.5- and 5-mm step sizes. Images were acquired at a rate of 146 frames/s. Acquired images were processed to automatically track the actual source using the correlation between a template image and each frame. The brachytherapy dose calculation formalism (AAPM TG43-U1) was applied to each frame to evaluate the transit dose contribution to the total dose. Results The differences in measured source positions from the nominal for dwell times equal to 0.1, 0.5, 1.0, and 2.0 s in treatment plans were approximately