Choice of reference axis in ocular wave-front aberration measurement.
Choice of reference axis in ocular wave-front aberration measurement.
复制标题
眼波前像差测量中参考轴的选择。
DOI:
--
复制
发表时间:
1998
影响因子:
1.9
通讯作者:
V. Lakshminarayanan
中科院分区:
文献类型:
--
作者:
C. Cui;V. Lakshminarayanan
The geometrical center of the pupil has often been used as the reference axis in ocular wave-front aberration measurement. However, the geometrical center of the pupil may shift when the pupil size changes under different conditions. In particular, for subjective methods, defining the center of the pupil precisely during the actual measurement is not always practical. Furthermore, the geometrical center of the pupil may not define the chief ray of the ocular optics because of the Stiles-Crawford apodization effect, which has a peak location that often deviates from the geometrical center of the pupil. We present the coefficient transformation table of the Taylor polynomial up to the sixth order with respect to reference axis shift. We illustrate the effect of wave-front aberration change with reference axis shift with experimental data. This type of wave-front aberration change could be a true measurement error if there is an error in defining the reference axis. We also propose using the coaxially sighted corneal reflex as a better reference axis in aberration measurement.