CE-MALDI interface based on inkjet technology
CE-MALDI interface based on inkjet technology
复制标题
DOI:
10.1002/elps.200900414
复制
发表时间:
2009-12-01
期刊:
影响因子:
2.9
通讯作者:
Dovichi, Norman J.
中科院分区:
文献类型:
--
作者:
Vannatta, Michael W.;Whitmore, Colin D.;Dovichi, Norman J.
An ink jet printer valve and a nozzle were used to deliver matrix and sample from an electrophoresis capillary onto a MALDI plate. The system was evaluated by the separation of a set of standard peptides. That separation generated up to 40 000 theoretical plates in less than 3 min. Detection limits were 500 amol for an ABI TOF-TOF instrument and 2 fmol for all ABI Q-TOF instrument Over 70% coverage was obtained for the tryptic digest of alpha-lactalbumin in less than 2 5 min.