Precision of flame atomic absorption measurements of arsenic, cadmium, calcium, copper, iron, magnesium, molybdenum, sodium, and zinc
Precision of flame atomic absorption measurements of arsenic, cadmium, calcium, copper, iron, magnesium, molybdenum, sodium, and zinc
复制标题
砷、镉、钙、铜、铁、镁、钼、钠和锌的火焰原子吸收测量精度
DOI:
10.1021/ac50012a017
复制
发表时间:
1977
期刊:
影响因子:
--
通讯作者:
J. Ingle
中科院分区:
文献类型:
--
作者:
N. W. Bower;J. Ingle
Procedures for evaluating the precision of atomic absorption measurements have been improved and applied to 8 elements. Measurements on a Varian AA-6 spectrophotometer indicate that there are many similarities in the precision characteristics of different elements. For instance, analyte absorption flicker noise limits the measurement precision of most elements at moderate absorbances. The actual value of the measurement precision at a given absorbance and the dependence of measurement precision on absorbance depend upon the calibration curve, the wavelength of analysis, the integration time, and the type of flame employed.