Sensitivity estimation in time-of-flight list-mode positron emission tomography.
Sensitivity estimation in time-of-flight list-mode positron emission tomography.
复制标题
DOI:
10.1118/1.4934374
复制
发表时间:
2015-11
期刊:
影响因子:
3.8
通讯作者:
J. Herraiz;Arkadiusz Sitek
中科院分区:
文献类型:
--
作者:
J. Herraiz;Arkadiusz Sitek
PURPOSE An accurate quantification of the images in positron emission tomography (PET) requires knowing the actual sensitivity at each voxel, which represents the probability that a positron emitted in that voxel is finally detected as a coincidence of two gamma rays in a pair of detectors in the PET scanner. This sensitivity depends on the characteristics of the acquisition, as it is affected by the attenuation of the annihilation gamma rays in the body, and possible variations of the sensitivity of the scanner detectors. In this work, the authors propose a new approach to handle time-of-flight (TOF) list-mode PET data, which allows performing either or both, a self-attenuation correction, and self-normalization correction based on emission data only. METHODS The authors derive the theory using a fully Bayesian statistical model of complete data. The authors perform an initial evaluation of algorithms derived from that theory and proposed in this work using numerical 2D list-mode simulations with different TOF resolutions and total number of detected coincidences. Effects of randoms and scatter are not simulated. RESULTS The authors found that proposed algorithms successfully correct for unknown attenuation and scanner normalization for simulated 2D list-mode TOF-PET data. CONCLUSIONS A new method is presented that can be used for corrections for attenuation and normalization (sensitivity) using TOF list-mode data.