Transparency and stability analysis of a surgical teleoperator system
Transparency and stability analysis of a surgical teleoperator system
复制标题
手术遥控操作系统的透明度和稳定性分析
DOI:
10.1109/haptic.2003.1191319
复制
发表时间:
2003
期刊:
影响因子:
--
通讯作者:
J. Wikander
中科院分区:
文献类型:
--
作者:
H. Flemmer;J. Wikander
This paper discusses transparency and stability issues for a three channel teleoperator system to be used in sensitive skullbase operations. Based on a design model of the combined master human system with parameter uncertainties and teleoperator optimal transparency conditions, the interaction (communication) channel controllers are derived. One of the interaction channel controllers is then altered to make the teleoperator closed loop meet requirements imposed by the robustness property passivity. Transparency of the teleoperator system is analysed by analysing the gain plots of all the four transfer functions in the H-matrix. The three channel teleoperator is found to have a bandwidth of around 90 rad/sec in the force reflection channel, a small influence on the reflected force from the master velocity for low frequencies, a bandwidth of 150 rad/sec in the forward channel and a slave which is relatively unaffected by external forces. The theoretical results from the stability analysis are verified by experimental results.