文献摘要

被引文献

相似文献

放射物理中心(RPC)现场剂量测定审查访问的一个关键组成部分是光子束校准,其中确定x射线的能量是一个关键因素。在任务组21方案中,将恒定SCD的20 cm和10 cm深度处10 cm x 10 cm场的电离比TPR 20/TPR 10用作射束质量的定量测量。RPC测量了大多数加速器型号和型号的TPR 20/TPR 10和在恒定SSD下685光子束(4-25 MV)的相应深度剂量百分比比(D-20/D-10)。TPR 20/TPR 10和D-20/D-10之间存在很强的相关性,这使得可以根据深度剂量百分比的测量来确定TPR比。对TG-21系数((L)/bar/p、P-壁、P-repl)中引入的不确定性进行分析,这些不确定性是由测量数据中的分布引起的,并转化为TPR比的测定结果,误差不显著(< 0.3%)。这种经验关系提供了一种替代技术,用于量化TG-21协议中定义的光束质量,而不会在输出校准中损失任何精度。在固定SSD上校准的人可能会发现这种技术是一种更容易和更快的方法。(C)1998年美国医学物理学家协会。
A key component of the Radiological Physics Center's (RPC) on-site dosimetry review visits are photon beam calibrations for which determination of the energy of the x rays is a key element. The ratio of ionizations, TPR20/TPR10, for a 10 cm x 10 cm field at depths of 20 and 10 cm for a constant SCD is used as a quantitative measure of beam quality in the Task Group 21 protocol. The RPC has measured both TPR20/TPR10 and the corresponding ratio of percent depth dose (D-20/D-10) at a constant SSD for 685 photon beams (4-25 MV) for most makes and models of accelerators. A strong correlation between TPR20/TPR10 and D-20/D-10 is presented which allows the determination of the TPR ratio from the measurement of the ratio of percent depth doses. An analysis of the uncertainty introduced in the TG-21 factors ((L) over bar/p, P-wall, P-repl) caused by the spread in the measured data and translated into the determination of the TPR ratio results in an insignificant error (< 0.3%). This empirical relationship provides an alternate technique for quantifying the beam quality defined in the TG-21 protocol without surrendering any loss of precision in output calibration. This technique may be found by those who calibrate at a fixed SSD to be an easier and quicker method. (C) 1998 American Association of Physicists in Medicine.