The spectrometer of the high-resolution multiposition Thomson scattering diagnostic for TJ-II
The spectrometer of the high-resolution multiposition Thomson scattering diagnostic for TJ-II
复制标题
TJ-II型高分辨率多位汤姆逊散射诊断光谱仪
DOI:
10.1016/s0920-3796(03)00387-9
复制
发表时间:
2003
影响因子:
1.7
通讯作者:
K. McCarthy
中科院分区:
文献类型:
--
作者:
J. Herranz;F. Castejón;I. Pastor;K. McCarthy
A high-resolution multi-position Thomson scattering system, which combines high accuracy and excellent spatial resolution, is operating on the stellarator TJ-II. Its diagnostic spectrometer is presented here. Its main characteristics, which provide the TJ-II Thomson scattering diagnostic with high spatial and spectral resolution, 2.2 mm and 2.1 nm, respectively, are described in this paper. In particular, the viewing optics has been designed in order to achieve low aberrations and distortions. The detector branch, which acts during a short pulse, ensures correct light intensification and synchronization with the laser beam. These attributes allow one to discriminate small-scale features that appear in density and temperature profiles in TJ-II plasmas, as well as steep gradients in both temperature and density profiles. Examples of results obtained demonstrating the characteristics of this system are presented.