Results of the Abbott RealTime HIV-1 Assay for Specimens Yielding "Target Not Detected" Results by the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test

Results of the Abbott RealTime HIV-1 Assay for Specimens Yielding "Target Not Detected" Results by the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test
复制标题

DOI:
10.1128/jcm.01394-09
复制
发表时间:
2010-03-01
影响因子:
9.4
通讯作者:
Yao, Joseph D. C.
Yao, Joseph D. C.
中科院分区:
医学2区
文献类型:
--
作者:
Babady, N. Esther;Germer, Jeffrey J.;Yao, Joseph D. C.

文献摘要

被引文献

相似文献

雅培实时HIV-1检测和COBAS AmpliPrep/COBAS TaqMan HIV-1检测(CTM)在来自美国所有9个地理区域的40个州的实验室获得的1190个独特的临床血浆样本中没有观察到明显的不一致结果,之前CTM产生“未检测到目标”的结果。
No significantly discordant results were observed between the Abbott RealTime HIV-1 assay and the COBAS AmpliPrep/COBAS TaqMan HIV-1 Test (CTM) among 1,190 unique clinical plasma specimens obtained from laboratories located in 40 states representing all nine U. S. geographic regions and previously yielding "target not detected" results by CTM.