Cone beam computed tomography artefacts around dental implants with different materials influencing the detection of peri-implant bone defects
Cone beam computed tomography artefacts around dental implants with different materials influencing the detection of peri-implant bone defects
复制标题
DOI:
10.1111/clr.13596
复制
发表时间:
2020-07-01
影响因子:
4.3
通讯作者:
Bornstein, Michael M.
中科院分区:
文献类型:
--
作者:
Schriber, Martina;Yeung, Andy Wai Kan;Bornstein, Michael M.
Objectives To investigate the diagnostic accuracy of cone beam computed tomography (CBCT) for the diagnosis of peri-implant bone defects of titanium (Ti), zirconium dioxide (ZrO2) or titanium-zirconium (Ti-Zr) alloy implants. Materials and Methods Ti, Ti-Zr or ZrO(2)implants with two diameters (3.3 mm, 4.1 mm) and one length (10 mm) were inserted in the angle of the mandible of six fresh defrosted pig jaws. Out of the 12 implants inserted, 6 served in the test group with standardized buccal peri-implant bone defects, whereas 6 served as control without bone defects. CBCTs were performed with three acquisition protocols (standard, high and low dose) using two devices. Four observers analysed CBCTs as follows: (a) presence of a peri-implant defect; (b) presence of peri-implant artefacts and impact on defect diagnosis; and (c) linear measurements of buccal peri-implant defect including height and width (in mm). Results CBCT device, CBCT settings, implant material, implant diameter and observer background did not significantly influence diagnostic accuracy. The sensitivity and specificity values were high for defect detection. ZrO(2)led to a lower than average diagnostic accuracy (0.781). The linear measurements of peri-implant defect were underestimated by