Effect of high-intensity focused ultrasound on whole blood with and without microbubble contrast agent
Effect of high-intensity focused ultrasound on whole blood with and without microbubble contrast agent
复制标题
DOI:
10.1016/s0301-5629(99)00043-5
复制
发表时间:
1999-07-01
影响因子:
2.9
通讯作者:
Crum, LA
中科院分区:
文献类型:
--
作者:
Poliachik, SL;Chandler, WL;Crum, LA
Using human whole blood samples with and without contrast agent (CA), we evaluated the effect of exposures to focused, continuous wave (CW) 1.1-MHz ultrasound for durations of 10 ms to 1 s at spatial average intensities of 560 to 2360 W/cm(2). Cavitation was monitored with a passive cavitation detector and hemolysis was determined with spectroscopy. In whole blood alone, no significant cavitation, heating or hemolysis was detected at any exposure condition. Conversely, cavitation and hemolysis, but not heating, were detected in whole blood with CA. A CA concentration as low as 0.28 mu L CA per mt whole blood at an intensity of 2360 W/cm(2) for 1 s resulted in measurable cavitation and a 6-fold increase in hemolysis compared to shams. Cavitation and hemolysis increased proportional to the concentration of CA and duration of exposure. In samples containing 4.2 mu L CA per mi, whole blood exposed for 1 s, a threshold was seen at 1750 W/cm(2) where cavitation and hemolysis increased IO-fold compared to exposures at lower intensities. HIFU exposure of whole blood containing CA leads to significant hemolysis in vitro and may lead to clinically significant hemolysis in vivo. (C) 1999 World Federation for Ultrasound in Medicine & Biology.