Time-resolved 3D MR velocity mapping at 3T: Improved navigator-gated assessment of vascular anatomy and blood flow
Time-resolved 3D MR velocity mapping at 3T: Improved navigator-gated assessment of vascular anatomy and blood flow
复制标题
DOI:
10.1002/jmri.20871
复制
发表时间:
2007-04-01
影响因子:
4.4
通讯作者:
Frydrychowicz, Alex
中科院分区:
文献类型:
--
作者:
Markl, Michael;Harloff, Andreas;Frydrychowicz, Alex
Purpose: To evaluate an improved image acquisition and data-processing strategy for assessing aortic vascular geometry and 3D blood flow at 3T.Materials and Methods: In a study with five normal volunteers and seven patients with known aortic pathology, prospectively ECG-gated cine three-dimensional (3D) MR velocity mapping with improved navigator gating, real-time adaptive k-space ordering and dynamic adjustment of the navigator acceptance criteria was performed. In addition to morphological information and three-directional blood flow velocities, phase-contrast (PC)-MRA images were derived from the same data set, which permitted 3D isosurface rendering of vascular boundaries in combination with visualization of blood-flow patterns.Results: Analysis of navigator performance and image quality revealed improved scan efficiencies of 63.6% +/- 10.5% and temporal resolution (