Accurate Measurement of Absolute Terahertz Power Using Broadband Calorimeter
Accurate Measurement of Absolute Terahertz Power Using Broadband Calorimeter
复制标题
DOI:
10.1007/s10762-018-0477-3
复制
发表时间:
2018-05-01
影响因子:
2.9
通讯作者:
Amemiya, Kuniaki
中科院分区:
文献类型:
--
作者:
Iida, Hitoshi;Kinoshita, Moto;Amemiya, Kuniaki
This paper presents a highly sensitive terahertz (THz) calorimeter developed using a magnetically loaded epoxy as a broadband absorber. The reflection loss of the absorber, which has a pyramidally textured surface, is less than 0.04, as determined using a THz time-domain spectrometer and a vector network analyzer. The THz calorimeter successfully enabled the measurement of the absolute THz power from a photomixer at microwatt levels at room temperature. The measurement uncertainties at a 95% confidence level were 6.2% for 13 mu W at 300 GHz and 5.6% for 1.5 mu W at 1 THz, respectively. Details of the evaluation and uncertainty analyses are also presented.