On-line detection of atmospheric formaldehyde by a conductometric biosensor.
On-line detection of atmospheric formaldehyde by a conductometric biosensor.
复制标题
DOI:
10.1016/j.bios.2006.03.018
复制
发表时间:
2007-01
影响因子:
12.6
通讯作者:
F. Vianello;Raffaella Boscolo-Chio;S. Signorini;A. Rigo
中科院分区:
文献类型:
--
作者:
F. Vianello;Raffaella Boscolo-Chio;S. Signorini;A. Rigo
Atmospheric formaldehyde (CH2O) was detected under continuous flow conditions by an on-line system comprising of a wet scrubber for a continuous transfer of the pollutant to an aqueous solution, a micro-reactor containing immobilized formaldehyde dehydrogenase (FDH) and a conductometric transducer. By this system atmospheric formaldehyde concentrations in the range 0.05–2ppm were detected with a sensitivity of 20μS/ppm. In this concentration range the immobilized enzyme oxidized all the sampled formaldehyde molecules to formic acid, avoiding cumbersome calibration procedures. The operational stability of the biosensor was at least 3 months, working continuously 10h/day at room temperature.