Cross-Calibration of Prodigy and Horizon A Densitometers and Precision of the Horizon A Densitometer
Cross-Calibration of Prodigy and Horizon A Densitometers and Precision of the Horizon A Densitometer
复制标题
DOI:
10.1016/j.jocd.2021.02.003
复制
发表时间:
2021-08-03
影响因子:
2.5
通讯作者:
Dawson-Hughes, Bess
中科院分区:
文献类型:
--
作者:
Reitshamer, Elise;Barrett, Kelsey;Dawson-Hughes, Bess
We performed this study to enable a reliable transition for clinical study participants and patients from a GE Lunar Prodigy to a Hologic Horizon A dual-energy X-ray absorptiometry (DXA) scanner and to assess the reproducibility of measurements made on the new DXA scanner. Forty-five older adults had one spine, hip, and total body scan on a Prodigy dual-energy X-ray absorptiometry (DXA) scanner and 2 spine, hip, and total body scans, with repositioning, on a new Hologic Horizon A DXA scanner. Linear regression models were used to derive cross calibration equations for each measure on the 2 scanners. Precision (group root-mean-square average coefficient of variation) of bone mineral density (BMD) of the total hip, femoral neck, and lumbar spine (L1-L4), and total body fat, bone, and lean mass, appendicular lean mass, and trabecular bone score (TBS) was assessed using the International Society of Clinical Densitometry's (ISCD's) Advanced Precision Calculation Tool. Correlation coefficients for the BMD and body composition measures on the 2 scanners ranged from 0.94 to 0.99 (p