Early Failure of Articular Surface Replacement XL Total Hip Arthroplasty
Early Failure of Articular Surface Replacement XL Total Hip Arthroplasty
复制标题
DOI:
10.1016/j.arth.2011.03.027
复制
发表时间:
2011-09-01
影响因子:
3.5
通讯作者:
Nadaud, Matthew C.
中科院分区:
文献类型:
--
作者:
Steele, Garen D.;Fehring, Thomas K.;Nadaud, Matthew C.
The ASR (articular surface replacement) XL (DePuy, Warsaw, Ind) metal-on-metal hip arthroplasty offers the advantage of stability and increased motion. However, an alarming number of early failures prompted the evaluation of patients treated with this system. A prospective study of patients who underwent arthroplasty with the ASR XL system was performed. Patients with 2-year follow-up or any revision were included. Failure rates, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores, and radiographs were evaluated. Ninety-five patients (105 hips) were included. There were 16 revisions. Thirteen (12%) were aseptic acetabular failures. Eight were revised for aseptic loosening; 4, for metallosis; 1, for malposition; 2, for infection; and 1, for periprosthetic fracture. Mean time to revision was 1.6 years (0.18-3.4 years). The ASR XL with a revision rate of 12% is the second reported 1 piece metal-on-metal system with a significant failure rate at early follow-up. This particular class of implants has inherent design flaws that lead to early failure.